Khron-Hite, 521

You are here: Calibrator » Voltage

Khron-Hite, 521
Above pictures can be different from actually sold item as well as the installed options. On request we'll send you high resolution pictures and options of the real item.
Product description
Technical details :

  • "Limit Of Error” (or Worst Case) methods.
    E Mode: ±(0.002% of setting + 0.0005% of range)
    I Mode: ±(0.005% of setting + 1µA) compliance with variable control)
  • Accuracies based on one full year calibration cycle and conservatively specified by using the "Limit-Of-Error" (or worst case methods).
  • 3 voltages ranges:
    ±100Vdc resolved to 100µV
    ±10Vdc resolved to 10µV
    ±100mVdc resolved to 0.1µV
  • 2 current ranges:
    ±100mAdc resolved to 0.1µA
    ±10mAdc resolved to 0.01µA
  • Programming: Integrated microprocessor, GPIB interface. Local and remote control.
  • The control logic is microprocessor based.
  • True bipolar control with balanced zero.A “crowbar” zero reference may be selected.
  • Resolution 6 digits or 1 ppm
VOLTAGE MODE

Range

Full Scale

Resolution

Current

Zo

100mVdc

±111.1110mVdc

100nVdc

EMF

20 ohms

10Vdc

±11.11110Vdc

10µVdc

100mAdc

10 milliohms

100Vdc

±111.1110Vdc

100µVdc

100mAdc

10 milliohms

1000Vdc

±1111.110Vdc

1mV

5mAdc

 

Noise and Ripple:

Range

0.1-100kHz

100mV

16µVrms

10V

50µVrms

100V

150µVrms

1000V

15mVrms


CURRENT MODE

Range

Full Scale

Resolution
1ppm

Voltage
Compliance**

10mAdc

±11.111 10mAdc

10nAdc

0-±100Vdc **

100mAdc

±111.111 0mAdc

100nAdc

0-±100Vdc **

**Compliance Voltage Limit Control: The Compliance voltage may be limited via a manual control. The limits are ±100V, ±46V, ±32V, ±18V, ±4V, ±1V.

Accuracy: (See definitions under voltage mode) ±[0.005% of setting + 1µA].
Stability (*non-additive): 8 hrs, ±0.001%; 90 days, ±0.0015%;1 year, ±0.002%.
Line & Load Regulation (*non-additive): ±0.0005% for a 10% line fluctuation and from short-circuit to full compliance.

Applications:
   Calibration of DVMs, DMM, meters, chart recorders, A/D converters, ATE, monitors, controllers, logging
   systems, etc.

  • Current: ± 0.01 μA - 10 mA dc
  • Type of calibration: Voltage - Current
  • Interface: GPIB
  • Voltage: ± 0.1 μV - 100V dc