Applications: This kit is used to calibrate the HP8753 network analyzer. Two terminations are supplied. This allows quick "Full" two-port calibrations. Additionally, with two terminations a critical triangle of agreement can be established between the two terminations. Simply calibrate a port using one termination and measure the other termination. If the other termination does not appear good then one or the other is not in specification and must be replaced. The kit can also be used to verify calibration and performance of other reflectometer type equipment such as the Anritsu SiteMaster or Bird Site Analyzers
Description:This kit contains two precision 50 ohm terminations. The termination is used during the "load" portion of a calibration. The DC accuracy of the termination is within 0.05 ohm.There is also an open. When installed this provides a dimentionally accurate outer transmission line for the female pin of the device being calibrated. Also provided is a precision short The short occurs within 0.1mm of the reference plane of the "N" connector. This is used to determine the open/short ratio before making your measurement.The components are contained in a durable plastic case. This not only protects the components during transit from one job to another, but also helps prevent loss.
Technical details :
Termination accuracy >-46 dB to 3.0 GHz
All components useable to DC - 3.0 GHz
The kit contains 2 termination 50Ω, 1 open and 1 short
Power rating: 2 Watts continuous
Calibration plot of termination included
Cerfitificate of calibration included