Main features:
· External voltage measurements up to 600V DC/AC
· Insulation resistance measurements with or without GUARD-technique up to 30 GOhm or 3 TOhm
· Nominal voltages between 50V and 1000V, fixed or variable in 10V steps
· Resistance measurements with DC-currents (Ik>200mA) and automatic polarity change up to 3 kOhm (100kOhm on analog display)
· Equivalent leakage current measurement (IPE) for checking electrical consumers ”after repairs” as per DIN VDE 0701
· Measurement of the local insulation resistance of floors or walls with optional probe.
· Permanently stored settings (e.g. limits)
Datasheet
- Insulation resistance 1Ω … 3 TΩ with measuring voltage 50 …1000 V DC with GUARD technology and auto-monitoring of prescribed limits
- Resistance 0.01 … 3 kΩ with automatic pole-changing and high short-circuit current as per DIN VDE 0413/4 (>200 mA)
- Equivalent leakage current as per DIN VDE 0701/1 at extra-low voltage, with automatic recomputation for mains-voltage level
- Standing surface insulation resistance 10Ω …30 MΩ at extra-low voltage sinusoidal 50 Hz
- DC or AC voltage up to 600 V