The 5200A Alternating Voltage Calibrator is a precision high-performance alternating voltage source with proven reliability and exceptional stability and accuracy over a broad frequency range. Wide output range & lab standard accuracy makes this instrument useful in both laboratory and manufacturing environments. Outputs from 100 µV RMS to 120 V RMS at a 50 mA current level. Frequency range from 10 Hz to 1.2 MHz with accuracy of ±(1.0% + 0.1%). Amplitude range is 1 mV to 1000 V in seven ranges. 6-digit resolution, yielding 1 nV steps on the 1 mV range increasing to 1 mV steps on the 1000 V range, with an error measurement provided in two ranges (±3% or ±0.3%). 20% overrange capability.
The Fluke 5200A has outputs from 100 microVrms to 120 Vrms at a 50 mA current level.
Frequency from 10 Hz to 1.2 MHz with accuracy of +/- (1.0% + 0.1%).
Amplitude range is 1 mV to 1000 V over seven ranges. ,
Features 6 digit resolution, yielding 1 nV steps on the 1 mV range
Fluke 5200A Programmable AC Calibrator
You are here: Calibrator » Voltage
Above pictures can be different from actually sold item as well as the installed options. On request we'll send you high resolution pictures and options of the real item.
Product description
Features
Voltage Ranges 1 mV, 10 mV, l00 mV, lV, 10V, 100V
Frequency Ranges l00 Hz, 1 kHz, 10 kHz, l00 kHz, 1 MHz
Amplitude Accuracy
10 Hz to 30 Hz
± (0.1 % of setting +0.005% of range)
1, 10, 100 volt ranges.
± (0.1 % of setting +10 jJ.V
1, 10, 100 millivolt ranges.
1, 10, 100 volt ranges.
± (0.1 % of setting +10 jJ.V
1, 10, 100 millivolt ranges.
30 Hz to 20 kHz
± (0.02% of setting +0.002% of range)
1, 10, 100 volt ranges.
± (0.02% of setting +10 jJ.V)
1, 10, 100 millivolt ranges.
1, 10, 100 volt ranges.
± (0.02% of setting +10 jJ.V)
1, 10, 100 millivolt ranges.
20 kHz to 100 kHz
± (0.05% of setting +0.005% of range)
1, 10, 100 volt ranges
± (0.05% of setting +20 jJ.V
1, 10, 100 millivolt ranges.
1, 10, 100 volt ranges
± (0.05% of setting +20 jJ.V
1, 10, 100 millivolt ranges.
0.1 MHz to 1 MHz
± (0.33% of setting +0.03% of range)
1, 10, 100 volt ranges.
± (0.33% of setting +30 jJ.V)
1, 10, 100 millivolt ranges.
1, 10, 100 volt ranges.
± (0.33% of setting +30 jJ.V)
1, 10, 100 millivolt ranges.
Frequency Accuracy
100-Hz to 100-kHz Ranges
±(1 % of setting
+0.1 % of range)
+0.1 % of range)
1-MHz Range
±(3% of setting
+0.3% of range)
+0.3% of range)
Max Output Voltage 120V rms, up to 83.33 kHz
Max Output Current 50 mA rms from 10% to 120% of range
- Current: 50 mA
- Frequency range: 1 MHz
- Type of calibration: AC Voltage
- Voltage: 100V
Price